Handy Digital Multimeters VOAC86A (Discontinued)Specifications 2

RECORD Mode

Sample Rate 50ms
Accuracy CREST mode where ±10 digits (±40 digits for AC signals) are added to the measurement function accuracy at 23°C ±5°C when the signal has fluctuated for 200ms or longer or the auto range has required 1.5 sec. or longer (except for the electrostatic capacitance and (DC + AC) voltage measurement).
Accuracy ±220 digits are added to the measurement function accuracy at 23°C ±5°C applied to signal fluctuation of 0.8ms or longer.

CREST Mode

Accuracy ±220 digits are added to the measurement function accuracy at 23°C ±5°C applied to the signal fluctuation of 0.8ms or longer.

SORT Mode

Sort Time approx. 0.2 sec. (except for electrostatic capacitance and (AC/DC) voltage measurement)
Accuracy ±5 digits are added to the measurement function accuracy at 23°C ±5°C.

Functions

  • 3.5- or 4.5-digit display
    • approx. 5 times per second when the display is 3.5-digit, 4,000
    • approx. 1.25 times per second when the display is 4.5-digit, 40,000
  • Hold
    • Function to hold measurement operations at all measurement functions
  • Manual or auto range
    • Manual range selection
    • Auto range selection
    • Note: These functions are canceled when the measurement range is manually switched in the RECORD, CREST, SORT, hold, recall, or relative mode.
  • Data store and recall
  • You can store all display data in the memory and recall it later. Saved data can be called from other functions.
  • Relative mode
  • Relative mode can be selected from three operation methods.
  • (1) D operation
    The relative value from the reference measurement value is displayed.
    The operation expression is "Display value = measurement value - reference measurement value".
    (2) Percent operation
    Measurement value is operated with respect to percentage to the reference measurement value and is displayed.
    The operation expression is "Display value = (measurement value - reference measurement value)/reference measurement value x 100 (%)".
    In this operation, the center of the bar graph is 0% and full scale display of ±200% or ±20% is used.
    (3) UNIT operation
    The rate of measurement value versus the reference measurement value is displayed as a positive integer.
    The operation expression is "Display value = measurement value/reference measurement value (UNIT)".
  • RECORD mode
    • The maximum and minimum measurement values are always saved in the internal memory.
      All sequential measurement values are calculated without the time restriction to derive the average value.
    • When the 40000 button is held for one second or longer, the symbol "R" appears on the LCD display section and the RECORD mode is set.
  • CREST mode
    • You can measure the peak value of an instantaneous signal that is 0.8ms wide.
    • When the 4000 button is held for one second or longer, the symbol "C" appears on the LCD display section and the CREST mode is set.
  • SORT mode
    • An alarm sounds when the measurement value is stable for about 0.2 second and the maximum, minimum, maximum minus minimum, and average values are calculated and recorded.
  • Backlight function
  • Alarm mute function
  • Selection of the line frequency from 50Hz or 60Hz (Note: Always set this value before measurement because this reduces the influence of hum noise from the commercial power supply frequency on measurement values.)
  • Auto power off (APO)
  • Input alert
  • RS mode (Data transfer only. Remote control function not available.)

General

Structure Dustproof construction
Operation Method Double integral method
Display
Display Section Main display section: 3.5-digit, 4,000 count or 4.5-digit, 40,000 count
  (However, the 5-digit, 99,999 count display is used for the frequency.)
  Sub display section: 4-digit, 9,999 count
Polarity Automatic selection
Update Speed 3.5 digits, 4,000 count display: approx. 5 times/sec.
4.5 digits, 40,000 count display: approx. 1.25 times/sec.
43 segment bar graph display: approx. 128 times/sec.
Low Battery Voltage Alert "" is displayed when the voltage drops to approximately 5.8V or lower.
Operation Temperature and Humidity 0°C to 35°C (80%RH or less), 35°C to 50°C (70%RH or less)
Storage Room Temperature -20°C to 60°C, 80%RH or less (excluding the battery)
Battery DC9V battery (IEC6LR61, JIS6F22, IEC6F22) x 1
Battery life: approx. 50 hrs. in the case of an alkaline battery (when used for DC voltage measurement)
Auto Power Off Available (on/off selection allowed. Power consumption: approx. 20mA)
Power Consumption approx. 12mA
Weight approx. 390g or 500g (including the holster)
Size approx. 87W x 35.5H x 186L mm
approx. 97W x 55H x 198L mm (including the holster)
Accessories
Test leads (with alligator clips and adaptors) One pair (red and black)
RS-232 adaptor (SC-521) 1
3.5" (1.44MB) floppy disk 2
Fuse (0.63A/500V IR 200kA) 1
Holster (attached) 1
Battery (attached)
This battery is for operation confirmation and may not have the specified life.
1