Home>Products>Universal Counters>SC-7207 / SC-7206 / SC-7205>Options

Universal Counters SC-7207 / SC-7206 / SC-7205 (Discontinued)Options

GP-IB interface (for SC-7205) SC-701

*Factory option
The SC-701 (standard equipment with the SC-7207 and SC-7206) that is built into the SC-7207, SC-7206, and SC-7205 universal counters can be used with the external GP-IB controller to remotely set remeasuring resets, measuring functions, time bases and calculations, and to externally output measuring results data.


Electrical and mechanical specifications are compliant with IEEE 488.1.
The following interface functions are available for GP-IB:
SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT1, C0
Equipped with REMOTE/LOCAL status indicator LED
Remote control items
Full remote control
(Except some functions including on/off for the power switch or remote control-related settings.)
Model
SC-7207 Equipped as standard
SC-7206
SC-7205 Optional (SC-701)

Address setting

The following items can be set from UTILITY - SYSTEM menu:
Any unit address can be selected from 31 types (0 to 30).
Either CR+LF or LF can be selected as the delimiter.

Digital I/O (for SC-7207, SC-7206 and SC-7205)
SC-702

* Factory option
The SC-7207, SC-7206 and SC-7205 universal counters with SC-702 installed are capable of controlling the start of measuring and outputting comparator calculation results (open collector). Part selection/inspection results can easily be checked visually by connecting an external lamp, etc.

Timing chart



Control signal

/TRIG (C-MOS input) Measuring is started by L pulse of 10 ms or more during /BUSY=Z period.
/TRIG signal is disabled during /BUSY=L period.
/BUSY (open collector output) During measuring: L (output transistor: on), during not measuring status: Z (output transistor: off)
/HI, /LO, /GO
(open collector output)
One position is L (output transistor: on) according to the comparator calculation results.

The rest is Z (output transistor: off)

Maximum rating of output terminal

•Voltage: DC50V
•Current: DC150mA
•Frequency response: DC to 1kHz

Maximum rating of input terminal

•Voltage: DC5V
•Frequency response: DC to 1kHz