IWATSU TEST INSTRUMENTS CORPORATION

  • Home
  • SiteMap
  • Products
  • Application
  • Download
  • Support
  • Company Profile
  • News
  • Contact Us
Products
  • Catalog
Products
Isolation Systems
  • System Configuration Example
  • Performance
  • High-voltage Probes
  • Model List

Home>Products>Isolation Systems DM-8000 and DM-8000H>Performance

Isolation Systems DM-8000 and DM-8000H Performance

Main Unit DM-8000 and DM-8000H

※When the acquisition control card (DM-610) is mounted.

Item
DM-8000
DM-8000H
Transceiver card connection
  Number of slots
6 slots (enabling connection of up to 12 isolation units)
Time axis
  Sweep range
1 ns/div to 20 s/div
  Clock accuracy
10ppm max
  Acquisition modes
Normal or Peak
Trigger system
  Mode
Auto, Normal, Single, Stop
  Sources
Up to 24 channels and an external trigger(EXT)
  Type

Edge or pulse width

  Trigger delay
Available
Interface
  Ethernet ports

Three 10BASE-T/100BASE-T ports

Three 1000BASE-T ports
Power supply unit
  AC power supply
100 V to 240 V (50/60 Hz)
Mechanical unit
  Dimensions and weight
351 (W) x 132 (H) x 420 (L) (mm), 6.9 kg ± 10%
  Operating temperature
0 to +40°C
  Performance guaranteed temperature
10 to +35°C
Accessories
  LAN cable
1
  Power supply cable
1
  Operating manual
CD-R (x1)

Isolation unit DM-9xx and DM-9xxL

Item

DM-900, DM-900L

DM-910, DM-910L
Signal input block
  Frequency Bandwidth
(-3 dB)
500 MHz
  Bandwidth limiter
20 MHz, 100 MHz
  Input impedance
1MΩ // 16pF
  Maximum input voltage
400 Vmax (DC + AC peak <= 5 kHz) CAT I
  Number of channels
2 (channels are not isolated)
1
  Input coupling
GND/ DC 1MΩ, AC 1MΩ
GND/ DC1MΩ
  Input sensitivity
2 mV/div to 10 V/div, 1-2-5 steps
CH1-MAIN:

50 mV/div to 5 V/div, 1-2-5 steps

CH2-ZOOM 2 mV/div to 1 V/div, 1-2-5 steps

  Offset range

2 mV/div to 50 mV/div: ± 1 V

100mV/div to 500mV/div : ±10V

1V/div to 10V/div : ±100V

CH1-MAIN:

50mV/div to 500mV/div:±10V

1V/div to 5V/div:±100V

CH2-ZOOM

2mV/div to 20mV/div:±2V

50mV/div to 1V/div:±20V

  DC gain accuracy
+/- (1.5% + 0.5% of ful-scale)
  Offset accuracy
Calculated by +/- (1.0% + 0.5% of full-scale + 1mV) at 10mV/div
  Probe sense
10:1, 100:1, 1000:1 (Auto detection/manual settings)
  Sample rate
1 GS/s (2 GS/s during interleave)
  Vertical axis resolution
8 bits
  Maximum memory length

DM-900:500k points per channel

DM-900L:16M points per channel

DM-910:500k points per channel

DM-910L:16M points per channel

Trigger system unit
  Trigger sources
CH1, CH2
CH1-MAIN
  Trigger slope
Positive / Negative
  Coupling
AC, DC, HFREJ, LFREJ
  Level range
125% of full scale
Interface
  Optical interface
One interface with 3 cables (optical fiber cable length: 2 to 200 m)
Power supply and battery unit
  Internal battery
3 battery packs (unit can operate on one battery)
  Battery charging
Can be charged by the main unit
  Battery charger
Option
  Output capacitance
285Wh
  Battery operation time
Approx. 12 hours of continuous operation (when using 3 batteries)
  Battery charging time
Approx. 6 hours
  AC power supply
100 V to 240 V (50/60 Hz)
Calibration signal
  Calibration signal
0.6 V/6 V (switchable)
Mechanical unit
  Dimensions (mm)
122.4 (W) x 258.4 (H) x 544 (L)
  Weight

7 kg ± 10% (excluding battery packs and accessories)
Battery pack weight: 660 g ± 10% per pack

  Operating temperature
0 to +40°C
  Performance guaranteed temperature
+10 to +35°C
Accessories
  Battery pack
3
  Power supply cable
1

Transceiver card DM-600

  Optical interface 2 interfaces (enabling connection of up to 2 isolation units)
  Operation indicator Status display via LED
  Interface Optical interface (2 interfaces with 3 optical fiber cables each)
  Mechanism Card inserted in main unit (DM-8000)
  Operating temperature 0 to +40°C
  Performance guaranteed temperature +10 to +35°C

IS Viewer DM-800


  ※IS Viewer is installed in the controlling computer (option) and is used to operate the isolation system and monitor waveforms.
  Operations +, -, ×, ÷, |×|, |÷|, ∫
  Parameter measurements Max, Min, p-p, Top, Base, Top-Base, RMS, CycleRMS, Mean, CycleMean, +/-Overshoot, TransitionTime, dv/dt, Freq, Period, +/-PulseCount, +/-PulseWidh, Duty, Integral, Integral(abs), Skew(%), Skew(Level)
  Other features Cursor, smoothing, channel deskewing, rescaling, offline viewer
  Saving waveforms CSV
  Saving images BMP, PNG, Clipboard
  Saving setup values With waveforms and without waveforms
  Controlling computer
  CPU Intel® Pentium®4 Processor or later
  RAM 2 GB or larger
  OS Windows® XP Professional SP2 / SP3
  Display At least WXGA (1280 x 768 pixels) recommended
(SXGA (1280 x 1024 pixels) is required for full display.)

Note 1. Intel and Pentium are registered trademarks or trademarks of Intel Corporation and its subsidiary companies in the United States of America and other countries.
Note 2. Windows is a registered trademark or trademark of Microsoft Corporation in the United States of America and other countries.
Top of this page
  • | Handling Personal Information |