Semiconductor Curve Tracer CS-3000 Series(UL approved)Specifications

Item CS-3100 CS-3200 CS-3300
Collector supply
  Mode High voltage AC, ± Rectified SIN, ± DC, ± LEAKAGE
Large current - Pulse only
  Maximum peak power

120 mW, 1.2 W, 12 W, 120 W, 390 W
*390 W can be selected excluding the maximum peak voltage

setting at 3,000 V range.

- Large current mode (400 W, 4 kW) Large current mode (400 W, 4 kW, 10 kW)
  High voltage mode
Maximum peak current
(maximum peak pulse current)
Maximum peak voltage
75 mA (150 mA) 3,000 V (2,500 V for AC)
750 mA (1.5 A) 300 V
7.5 A (15 A) 30 V
  Loop correction The stray capacitance between the collector terminal and the ground of the fixture is compensated by the hardware in the high voltage mode. Digital compensation is also available.
  Large current mode (pulse only) -
Maximum peak current Maximum peak voltage
400 A 40 V
40 A 40 V
Maximum peak current Maximum peak voltage
1,000 A 40 V
400 A 40 V
40 A 40 V
  Pulse width/measurement point - The pulse width is variable in the range from 50μs to 400 μs.
Measurement points can be specified (with a resolution of 10 μs/step).
  Maximum number of data points 20 to 1,000 points can be specified per trace.
Step generator
  Current mode Amplitude range: 50 nA to 200 mA, 21 ranges, at 1-2-5 step
Maximum current: ±2A / Offset: ±10 times STEP AMPLITUDE setting
  Voltage mode Amplitude range: 50 mV to 2 V, 6 ranges, at 1-2-5 step
Maximum voltage:±40V / Offset: ±10 times STEP AMPLITUDE setting
  Step rate Staircase wave: 2times of 50 Hz or 60 Hz
(50 Hz or 60 Hz in the AC mode)
Staircase wave: 2times of 50 Hz or 60 Hz (50 Hz or 60 Hz in the AC mode)
Pulse: The pulse changes in the range from 80ms to 1,000 ms.
(The lowest frequency limited by the maximum peak power consumption setting.)
  Pulse step The pulse width is variable in the range from 50 μs to 400 μs at 10μs resolution.
  Number of steps 0 to 20 steps
AUX output
  Range OFF, -40V to +40 V, variable with at 100mV resolution
Vertical axis
  Collector current Range High voltage mode: 1 μA/div to 2 A/div, 20 ranges, at 1-2-5 step.
- High current mode: 100 mA/div to 50 A/div, 9 renges, at 1-2-5 step.

High current mode: 100 mA/div to 100 A/div, 10 renges, at 1-2-5 step.

Accuracy 2% of readout + 0.05 × VERT/div setting value or lower (The internal loop correction error below must be added to the left formula.)
3 kV range: 6 μA, 300 V range: 1 μA, 30 V range: 0.5 μA (Defined only for the area of 10% or more of each voltage range.)
  Emitter current (LEAKAGE) Range 1 nA/div - 2 mA/div, 20 ranges,at 1-2-5 step
Accuracy 2% of readout + 0.05 × VERT/div setting + 1 nA or lower
Horizontal axis
  Collector voltage Range High voltage mode: 50 mV/div to 500 V/div, 13 ranges, at 1-2-5 step
- High current mode: 50 mV/div to 5 V/div, 7 ranges, at 1-2-5 switching
Accuracy 2% of readout + 0.05 × HORIZ/div setting or better
  Base/emitter voltage Range 50 mV/div to 5 V/div, 7 ranges, at 1-2-5 step
Accuracy 2% of readout + 0.05 × HORIZ/div setting or better
Miscellaneous
  Monitor 8.4-inch color TFT-LCD (SVGA 800 x 600 pixels)
  Data saving/reading Internal: Memory (setup: 256, REF waveform: 4)
External: Removable storage connected to the USB port
(setup, waveform saving / recall, screen hard copy)
  USB 1 port (USB1.1)
  Remote control Remote control by LAN : 1 port (100BASE-TX)
  Power supply / power consumption Power supply input range: 100 to 240 V AC,
50/60 Hz Power consumption: 500 VA
  Accessories CS-301 (fixture S),
CS-500 (test adapter),
operation manual,
power cord
CS-302 (fixture M),
CS-500 (test adapter),
wire set,
operation manual,
power cord
Mechanism section
  Dimensions (mm) 424 (W) x 555.2 (L) x 221 (H) (without the external projection portion) 424 (W) x 555.2 (L) x 354.5 (H)
(without the external projection portion)
  Weight Approx. 30 kg (without optional accessories) Approx. 45 kg (without optional accessories)
  Environmental conditions Operating temperature: 0 to +40 °C,
performance guarantee temperature: +10 to +35 °C