| Item |
SS-250 |
SS-240A |
| Bandwidth |
DC to 100 MHz
(-3dB) |
DC to 50 MHz
(-3dB) |
| Maximum input current |
Maximum input range |
30 A rms |
30 A rms |
| Maximum peak current |
50 A peak, non-continuous |
50 A peak, non-continuous
50 A peak at pulse width <= 10 µs |
| Output sensitivity |
0.1 V/A |
| Sensitivity accuracy |
±1.0% of reading ±10 mA for probes only (in the range from 0 to 30 A rms, DC, and AC 45 to 66 Hz)
±2.0% of reading for probes only (50 A peak or less and over 30 A rms, DC, and in the range of AC 45 to 66 Hz) |
| Noise |
2.5 mA rms or less (observation
using an oscilloscope with a 20 MHz band) |
| Power consumption |
5.3 VA max. |
5.6 VA max. |
| Power supply voltage |
±12 V ±0.5 V |
| Operating temperature and humidity |
0°C to +40°C, 80% RH or
less
(no condensation) |
| Storage temperature and humidity |
-10°C to +50°C, 80% RH
or less
(no condensation) |
| Operating elevation |
2000 m or less (air pressure approx.
79 kPa), indoor |
| Effects of external magnetic field |
5 mA max. or equivalent (in DC, 60 Hz, 400
A/m AC magnetic field) |
20 mA max. or equivalent (in DC, 60 Hz, 400
A/m AC magnetic field) |
| Maximum usable circuit voltage |
300 V, CAT I (insulating conductor) |
| Conductor diameter that can be measured |
ø 5 mm |
| Cable length |
Sensor cable |
Approx. 1.5 m |
| Power cable |
Approx. 1 m |
| Size |
Sensor part |
Approx. 175 (W) x 18 (H) x 40 (D)
mm |
| Termination part |
Approx. 27 (W) x 55 (H) x 18 (D)
mm |
| Weight |
Approx. 240 g |
Approx. 230 g |
| Complied with standards |
Safety |
EN61010-2-032:2002
Measurement category I (expected overvoltage: 1500 V),
Pollution degree: 2 |
| EMC |
EN61326:1997+A1:1988+A2:2001 |
| Accessories |
Instruction manual (1),
hard case (1) |
Instruction manual (1),
soft case (1) |